Specifications - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1

Specifications

Dimensions
Weight
Electrical Requirements
Power supply
Rated voltage range and frequencies 100 to 240 VAC 50/60 Hz
Stability
Power consumption
Fuses
Grounding
Interruptability
Battery
Battery type
Battery charging time
Battery operation
Recharging
Battery operaton alarm
early warning
final warning
Battery state indicator
Environmental Requirements
Operating temperature
Storage temperature
Atmospheric pressure
Humidity
In-built recorder
Power consumption
Principle
Print resolution
Vertical
Horizontal
Paper width
Traces
Print speed
2
.
Document no
M1144958-001
335 x 205 x 340 mm /13.2 x 8.1 x 13.4in
(W x D x H)
8.5 kg (19.4 lbs) (w/o modules)
(limits 90-264 VAC)
±10 % of nominal voltage
140 VA (max)
2x2A, T/SLOW, 250 V
Hospital grade
Data memory and alarm settings are saved up to 15 minutes
after monitor standby
NiMH, 3.5Ah
10 hours to full capacity
up to 1.5 hours typical on full charge at 23 °C / 73 °F
when connected to mains power
'Battery low' note
'Battery low' alarm
Capacity bar displayed on the screen
10 - 35 °C / 50 - 95 °F
-10 - 50 °C / 14 - 122 °F
500 - 800 mmHg (660 - 1060 mbar, 66 - 106kPa)
10 - 90 % non-condensing
3 W
Thermal array
8 dots/mm (200 dots/inch)
32 dots/mm (800 dots/inch) at a speed of 25 mm/s and slower
50 mm, printing width 48 mm
Selectable 1, 2, or 3 traces
1, 6.25, 12.5, 25 mm/s

Advertisement