Before Beginning; Visual Inspection - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors

3.2.3 Before beginning

Make sure that the monitor is switched to standby.
Disconnect the power cord and press the service reset switch until the LED next to it turns
off.
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net
cable. Take out any memory cards, if installed.
Remove the AC/DC Power Unit.
Disassemble the Side Frame Unit.
Detach the battery cables from the Power Management Board.

3.2.4 Visual inspection

1.
Side Frame Unit
Check the Side Frame Unit:
2.
PMB and the Connector Boards
Check the Frame Unit from inside. Check the PMB and the Connector Boards:
3.
Frame Unit reattached
Reconnect the battery cable.
Reattach the Side Frame Unit.
Reattach the Frame Unit back plate.
Reattach the AC/DC power unit.
NOTE: Make sure that all insulation plates are placed correctly before reattaching the rear
cover.
4.
Battery slot
Internal parts under bottom cover
Detach the bottom cover and check internal parts:
12
.
Document no
M1144958-001
screws are tightened properly
cables are connected properly
connectors are clean and intact
there are no loose objects inside the Frame Unit
all screws are tightened properly
all connectors are clean and intact
cables are connected properly
the battery cable is connected properly