Disassembly And Reassembly; To Disassemble The Display Unit F-Cm(C)1 Rev.04 And 05 - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

3.3 Disassembly and reassembly

WARNING
The S/5 Compact Anesthesia/ Critical Care Monitor is always energized by
the internal battery. A short circuit may cause internal damage. Do not
touch any exposed wiring or conductive surface inside, this may cause an
electric shock.
CAUTION
Perform leakage current measurement whenever service or repair has been
done in the monitor.
WARNING
The backlight circuit runs on a high voltage. Do not touch the inverter board
or the backlight tube leads when powered.

3.3.1 To disassemble the Display Unit F-CM(C)1 rev.04 and 05

1.
Set the Compact Monitor lying on its rear side.
2.
Turn the quick lock screw at the bottom of the Display Unit counter clockwise to release it.
3.
Pull the Display Unit carefully upwards.
4.
Set the Display Unit lying on its front side; be careful not to damage the ComWheel or the
LCD screen.
5.
Open and detach the four screws with the washers at the corners of the Display Unit.
NOTE:
X7
X5
Display Unit F-CM(C)1 rev. 04
When reassembling the cover, be careful not to pinch the Network cable.
Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
Document no. M1144956-003
33