Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Switch the monitor to standby and disconnect the power cord. Wait for two minutes, then
reconnect the power cord and switch the monitor back on.
Check that the trend information is still available.
53. Service reset switch check
Switch the monitor to standby and disconnect the power cord. Press the service reset
switch for at least five seconds. Check that the LED next to the switch turns off, and also
the fan turns off.
Reconnect the power cord and turn the monitor back on. Check that all trends have now
been erased.
54. Battery capacity
Test the batteries by performing the following:
Replace the batteries, if the time was less than 45 minutes.
55. Service Log reset
Enter the Service Log in the service menu:
Monitor Setup
Clear the content of the Service Log by selecting RESET LOG from the menu.
56. Electrical safety check
Perform an electrical safety check and a leakage current test.
Check that the monitor functions normally after the performed electrical safety check.
57. Final cleaning
Switch the monitor to standby, disconnect the power cord and perform final cleaning.
Fill in all necessary documents.
32
.
Document no
M1144956-003
detach all modules
keep the monitor connected to the mains power until the batteries are fully charged
install a Hemodynamic Module and disconnect the power cord
notice the time the monitor stays on with batteries
Install Service (Password 16-4-34)
Service (Password 26-23-8)
Service Log