8

Remote Access

Remote Access menu is for setting the interface with
an external PC that runs the Virtual Support software,
which is a service diagnostic tool for remote
diagnostics purposes.
When the access mode is set to Normal (the default),
the PC can only read service menu data from the
monitor.
With the access mode Extended, the PC can also
activate the monitor's service menu functions.
NOTE: Patient monitoring is not possible, when the
access mode Extended is selected. The access mode
turns back to Normal, if the monitor is restarted or if
the service menu is closed.
82
.