Display; Keyboard(S); Display Unit; Frame Unit - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

2.2.2 Display

1.

2.2.3 Keyboard(s)

1.
Tests with the Command Board:
Tests with the ARK Keyboard:
Tests with the Remote Controller:

2.2.4 Display Unit

1.
Check that the clock on the screen shows correct time.
Readjust the time and date, if necessary.
2.
Enter the Service Log service menu.
Check the content of the Service Log for possible problems.

2.2.5 Frame Unit

1.
Check that the Frame unit fan is running.

2.2.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX

1.
Check that the fan is running.
2.
If the module contains membrane keys on the front panel, press each of the keys for at
least one second and check that they are identified.
NOTE: The Compact Airway Module membrane keys require S/5 main software.
3.
Wait until the message 'Calibrating gas sensor' disappears from the screen, then enter
the Compact Airway Module General service menu.
Check that the picture on the screen is adjusted correctly..
Help
Press the
key. Turn the ComWheel in both directions and check that the cursor
in the menu moves correspondingly. Select Normal Screen and check that the
menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Enter the Keyboard service menu
Check the function of the ComWheel.
Press all keys, except
from the loudspeaker, or the 'Message count' value in the service menu increases.
Modify
Press the
on the monitor screen.
Enter the Keyboard service menu.
Check the function of the Comwheel.
Press all keys. Check that each key produces a sound from the loudspeaker, or the
Message count value in the service menu increases.
Modify
Print
and
. Check that each key produces a sound
Print
and
keys and check that the corresponding menus appear
Document no. M1144953-001
13