MAINTENANCE

The Everpure
Conserv
HE-III System requires very little
®
®
maintenance. Regular cartridge replacement, system
recovery verification, and sanitization are the only normal
maintenance requirements. Typical service schedules are as
listed below.
Every 12 Months (Recommended)
Replace all of the cartridges and sanitize the permeate
storage tank.
Switch and valve starting positions:
• Power - OFF
• Soft Water Inlet Valve - CLOSED
• Hard Water Inlet Valve - CLOSED
• Blended Water Outlet Valve - CLOSED
1. Remove the cover on the Conserv HE-3 RO System.
NOTE: When the cartridges are removed, water will leak
from the heads. Be prepared to catch or clean up any
spills.
2. OPEN the outlet valve. Use the application equipment to
release water pressure. CLOSE the outlet valve.
3. Have the new replacement cartridges ready to install.
Remove all the old cartridges by grasping the outside of
the cartridge. Push up and turn it to the left until it stops.
The cartridge will then pull straight down and out.
4. Refer to "Prefilter and RO Cartridge Installation and
System Start-up" on page 12. Perform Steps 1
through 14.
5. Refer to the section "
page 14. Perform Steps 2 through 17.
10
6. Refer to the section "
Cartridge" on page 16, Perform Steps 1 through 10.
The system is ready to run. Use application blended water
for five (5) minutes and test the TDS. If the measured
TDS reading is not within the required specification of the
application, then the blend should be adjusted.
7. To adjust the TDS of the blended application water, refer
to the "
Adjust The Blending Valves" on page 16.
11
35
EVERPURE
CONSERV
HE-3 RO SYSTEM INSTALLATION AND OPERATION GUIDE
®
®
7
Prepare Storage Tank" on
Flush Everpure Postfilter

COMPLIANCE STATEMENTS

Check for compliance with state and local laws and regulations.
Do not use with water that is microbiologically unsafe, or of
unknown quality without adequate disinfection before or after
the system.
Substances reduced are not necessarily in your water. System
must be maintained according to manufacturer's instruction,
including replacement of filter cartridges.
This system contains a replaceable mineral reduction (RO)
module critical for the effective reduction of total
dissolved solids.
Blended water should be tested periodically to verify that the
system is working properly.
This reverse osmosis system contains a replaceable
component critical to the efficiency of the system.
Replacement of the reverse osmosis component should
be with one of identical specifications, as defined by the
manufacturer, to assure the same efficiency and containment
reduction performance.
Recovery rating refers to the percentage of the influent water to
the membrane portion of the system that is available to the user
as reverse osmosis treated water when the system is operated
without a storage tank or when the storage tank is bypassed.
Efficiency rating means the percentage of the influent water
to the system that is available to the user as reverse osmosis
treated water under operating conditions that approximate
typical daily usage.
Pentair
Everpure
Technical Support:
®
®
Tel: 800.942.1153 (US Only) • 630.307.3000 Main •
Email: servicespecialist@pentair.com
System Tested and Certified by NSF International against NSF/ANSI Standards 42, 53 and 58 for the reduction of:
Standard No. 42 — Aesthetic Effects
Standard No. 53 — Health
Effects
Chemical Reduction
Mechanical Filtration
Taste and Odor
Chlorine
Chloramines
Mechanical Filtration
Nominal Particulate Class I
Water Treatment
Applicance - E359339
Standard No. 58
Mechanical Filtration
Turbidity
Total Dissolved Solids
Cyst
(TDS)
35