6510 Signal Input Module - Emerson Machinery Health Expert CSI 2600 User Manual

Table of Contents

Advertisement

Signals from 4-20 mA devices, RTDs, thermocouples, and other specialized sensors require
external conditioning electronics that convert the sensor signal into a voltage signal the
CSI 2600 can accept on its BNC input panel. The CSI 2600 can supply power to
piezoelectric sensors from the BNC connection.
The CSI 2600 has 4 BNCs for tachometer inputs on the rear panel. The tachometer BNC
inputs accept voltage signals only and do not furnish tachometer power.
When the CSI 2600 is in transient data collection mode, not all signals connected to a CSI
2600 need to be designated as transient. But since the Transient system is used to detect
unexpected events using its continuous monitoring capability, it's best to designate every
channel to be transient to make sure a special event is not missed.
4.8

6510 Signal Input module

The 6510 Signal Input module combines the features of Signal Input, Tachometer Input,
and I/O Relays to allow a combination of sensor and relay types in one module.
The 6510 provides 12 channels of vibration or process sensor inputs, 2 channels of
tachometer sensor inputs, and 2 optically-isolated I/O relay channels.
See
Section A.2.2
MHM-97453, Rev 1
for specific calibration information.
Hardware configuration: overview
29

Advertisement

Table of Contents
loading

Table of Contents