D.4

Rear terminal power connections

Figure D-2:
D.5

CSI 2600 DIP switch settings

Each signal input channel has an associated DIP switch for connecting accelerometer
power. For accelerometer channels that require power from the CSI 2600, set the
associated DIP switch to the right, or On position. For sensor channels that do not require
power, set the associated DIP switch to the left or Off position.
Note
For the Tach and Relay channels, the DIP switches must be left in the Off position.