Consider Housing Rotation; Connect The Power Module - Emerson Rosemount 3051CF DP Quick Start Manual

Pressure transmitter and flowmeters
Table of Contents

Advertisement

Quick Start Guide
Note
A weep hole has been designed into the transmitter for safety and leak detection. If fluid
begins to leak from the weep hole, isolate the process pressure, disconnect the
transmitter, and reseal until the leak is resolved.
2.2

Consider housing rotation

To improve visibility of optional LCD display:
1. Loosen the housing rotation screw.
2. First rotate the housing clockwise to the desired location. If the desired
location cannot be achieved due to thread limit, rotate the housing counter
clockwise to the desired location (up to 360° from thread limit).
3. Re-tighten the housing rotation screw (see
Figure 5. Housing Rotation
A. Housing rotation screw (
2.3

Connect the power module

1. Remove the power module cover.
2. Connect the green power module (see
Figure 6. Power Module
A. Power module
8
A
5
/
-in. hex wrench required)
64
Figure
5).
Figure
6).
A
June 2017