Software configuration
Procedure
1.
2.
52
Note
Do not add a CSI 2600 to an existing ethernet network until its IP addresses (CPU board,
Transient board) have been verified and changed, if necessary, to be compatible with
addresses already in use on the existing network.
Turn on the CSI 2600 and start a terminal session.
A screen similar to the following will appear during the boot process:
Figure 5-3:
HyperTerminal System Boot screen
When the boot process has completed, type bootChange and press Enter to configure
the unit. This command is case-sensitive.
A list of boot parameters appears one line at a time. When configuring the Main
Processor, the screen will look similar to this:
Figure 5-4:
Main Processor boot parameters
When configuring a Transient Processor, the screen will look similar to this (note the
difference in the file name value, which includes a "t" for Transient, and the different
boot flags:
MHM-97453, Rev 1