GS recalibration should be performed:
Product Support personnel can guide a plant engineer or technician through GS calibration
over the phone.
B.1.4
Digital Signal Processor (DSP) calibration
Digital Signal Processor (DSP) calibration:
DSP recalibration should be performed:
Product Support personnel can guide a plant engineer or technician through DSP
calibration over the phone.
B.1.5
Transient calibration
The Transient unit includes two processing boards—a Main Processor board and a
Transient board. Both boards include separate Digital Signal Processors. The DSP on the
Transient board uses an internal calibration table, in much the same way that the Main
Processor DSP circuit does. However, it is calibrated separately, and not as part of a DSP
calibration for a Main Processor.
Transient calibration:
MHM-97453, Rev 1
uses a special calibration utility program (DHM).
annually.
whenever a 6510 Signal Input module is replaced.
whenever a 6560 Processor module is replaced.
if the calibration table has a status of "Unknown".
uses a CPU board's TSG output signal.
does not require that any wire harnesses be disconnected.
is completed in about 30-40 minutes.
does not require any special cables or test equipment.
uses a special calibration utility program (DHM).
annually.
whenever a 6510 Signal Input module is replaced.
whenever a 6560 Processor module is replaced.
if the calibration table has a status of "Unknown".
uses a Processor module's TSG output signal.
does not require that any wire harnesses be disconnected.
is completed in about 30-40 minutes.
does not require any special cables or test equipment.
uses a special calibration utility program (DHM).
System calibration
85