Sensica™ UO System for ICU Instructions for Use, Model No. SCCS-1001, ver. 1001
Table 13. Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the Sensica™ UO System for ICU - for Medical
Electrical Equipment and Medical Electrical Systems that are not Life-Supporting
The Sensica™ UO System for ICU is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Sensica UO
System for ICU can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
Sensica UO System for ICU as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
P/N 005-0175 Rev A
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
Adaptec Medical Devices
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23
67