Sensica™ UO System for ICU Instructions for Use, Model No. SCCS-1001, ver. 1001
C
16: A
S
HAPTER
LERTS
ETTINGS
Touch the "Alerts" button to access the Alerts Setting Screen. This screen allows personalization for
the following alerts: UO low, UO high, a sudden increase in UO, and urine bag full. It also allows a
selection for the catheter use time display reminder and operator PIN requirement.
After making desired selections, touch "Save and Return." This will save the changed data and
return the screen to the previous screen accessed. Touching the "Return" button will return the
screen to the previous screen accessed without saving any changes to the data. To reset the Alert
Settings to original selections, touch the "Reset to Factory Defaults" button.
Figure 48: Administrator Settings - Alert Settings Screen
16.1 Urine Output Low Alert Setting
Select the amount of urine output (ml) over a period of time (hours) to alert the operator that the UO
is low. The default setting is less than 50ml urine output over a two-hour period (i.e., less than
25ml/hr average for two hours). Selection values for UO are: 5, 10, 15, 20, 30, 40 and 50 mls.
Selection values for hours are: 1, 2, 3 and 4 hours.
16.2 Urine Output High Alert Setting
Select the amount of urine output (ml) over a period of time (hours) to alert the operator that the UO
is high. The default setting is greater than 400ml urine output over a two-hour period (i.e., more
than 200ml/hr average for two hours). Selection values for UO are: 100, 200 ,300, and 400 mls.
Selection values for hours are: 1, 2, 3 and 4 hours.