Sensica™ UO System for ICU Instructions for Use, Model No. SCCS-1001, ver. 1001
Figure 6: Clockwise twist-lock connection of Sensica™ UO Ring to stand
The Sensica UO ring will provide continuous operation for 336 hours (14 days) when connected to
the Sensica UO System for ICU. If the patient requires longer monitoring than 14 days, a new ring
can be connected to the stand without interrupting data collection or patient history.
1.1.3 Sensica™ UO Patient Kit: Sensica™ UO Tube Band with Measuring Strip
An L-shaped, Sensica UO Tube Band with Measuring Strip is included with the patient kit. This is
used to put a band around the drainage tubing at the correct distance for tethering the tubing to the
tube holder. Once the appropriate length of tubing is measured, the peel-off backing on the band
can be removed to expose glue and the band wrapped around the drainage tubing, creating a
permanent tube band. Then, the perforated measuring strip can be torn off and discarded.
To properly insert the drainage tubing in the tube holder, the tube band is aligned with the tube
holder "window" and the tubing is squeezed into the holder for a secure fit. This ensures accurate,
ongoing UO measurement data. The tube band must be in the tube holder window and the tubing
securely settled into the holder for the system to function correctly.
P/N 005-0175 Rev A
Adaptec Medical Devices
Figure 7: Using Sensica™ UO Tube
Band to place tubing in Sensica UO tube
holder
12