Emissions Test
RF Emission CISPR 11
RF Emission CISPR 11
Harmonic Emissions IEC
61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Immunity test
ESD IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-
11
Power Frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
Compliance
Group 1
Class B
Class B
Class B
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV differential mode
± 2 kV common mode
0% 0.5 periods 0°
40% 5 periods 0°
70% 25 periods 0°
0% 250 periods 0°
3 A/m
Page 63/68
Electromagnetic environment - Guidance
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The device is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low
voltage power supply network that supplies
buildings used for domestic purposes.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV differential mode
± 2 kV common mode
0% 0.5 periods 0°
40% 5 periods 0°
70% 25 periods 0°
0% 250 periods 0°
3 A/m
Electromagnetic
environment -
Guidance
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material the relative
humidity should be at
least 30 %.
Mains power supply
should be that of a
typical commercial or
hospital environment.
Mains power supply
should be that of a
typical commercial or
hospital environment.
Mains power supply
should be that of a
typical commercial or
hospital environment.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.