Review video screen
Touch Play button (1) in navigation bar to play back the entire measurement video in slow
motion. During playback Play button (1) is replaced by Pause button (2). Touch Pause button
(2) to interrupt playback and inspect a single picture closely. The inspection of images is
primarily used for detecting media opacities that do not cause a measurement to abort. Touch
Play button (1) to continue playback.
Note:
Check after every measurement whether both pupils are equally bright and
whether structures are visible within the pupils. If brightness is different or
structures can be seen, the eyes need to be examined closely.
Video pictures can help identifying the reason for a measurement to abort. Pay attention to
both pupils. They have to be visible entirely and must not be covered or partly covered by hair,
eyelashes or eyelids for example.
Screenshot 6: Review video screen
1)
2)
Page 23/68
Need help?
Do you have a question about the A12C and is the answer not in the manual?