(clear entry) key clears any parameter in the active field. Pressing
diagnostic messages. If a diagnostic message does not clear when
Troubleshooting.

System Keys

START
STOP
ENTER
Specifications
Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.
Specification
Speed
Time
1-8
Pressing ENTER then START begins the run. (START must be pressed within 5 seconds of
pressing ENTER.) This key combination can also be used to abort a deceleration process and
restart the centrifuge.
Pressed to end a run according to the deceleration setting. The red light on the STOP key
flashes while the rotor decelerates. An audible tone sounds when the rotor comes to a
complete stop. Deceleration can be terminated and the centrifuge restarted by pressing
ENTER then START.
Pressed to save parameters to system memory, and to start the centrifuge. ENTER, then
START must be pressed to start the centrifuge).
Description
• Set speed
— 100 to 10,000 RPM (ARIES rotors: 1000 to 10,000 RPM)
in 10-RPM increments; or RCF from 200  g to maximum
for the installed rotor (in 10  g increments)
• Speed control
— actual rotor speed 25 RPM; g force control sets speed
to the nearest 50 RPM
• Set time
— 1 min to 99 hr 59 min (timed run), total centrifugal effect
(g-seconds), or continuous (hold)
• Time display
— indicates run time remaining (timed run), g-seconds
setting and time remaining to accumulate set g seconds
(g-seconds run), or elapsed time (hold run)
CE
will also clear some
CE
is pressed, see
CHAPTER 3,
PN J3HC-IM-5AD