Beckman Coulter Avanti J-HC Instructions For Use Manual page 42

High-capacity centrifuge
Hide thumbs Also See for Avanti J-HC:
Table of Contents

Advertisement

Operation
Programmed Operation
b. If you enter a number that is already assigned to a program, you can either overwrite the
program or select another program number.
3
Enter run parameters (rotor, speed, time, temperature, acceleration/deceleration settings)
for step 1.
The message "
a. Press
NOTE
To program a delayed start, set the time in step 1 to the required delay length and set the speed
in step 1 to 0 RPM. Enter run parameters in step 2, setting the step 2 time to any value greater than
00:00. The run will start when the step 1 set time has elapsed.
4
If you want to program one step only, press the
program mode.
The parameters are saved to step 1.
If you are overwriting a program that contains user-entered settings in step 2, those values are
retained until you change them.
a. To check step 2 settings, press the
5
To add a second step to program 1, press the
6
Enter parameters for step 2.
All parameters for step 1 and 2 are saved under the selected program number.
a. Press
2-14
Rotor : ---
Speed : ---
RPM
=
Time :
00:00
Temp : 20
Max Temp : 24 -
Accel : MAX
Decel : MAX
Prog Recall:
Setup: 1
Enter parameters for the selected step
Switch steps, save, or exit program mode
ENTER
when programming is complete.
Rotor : JS-5.0
Speed : 5,000
RPM
= 7,480
Time : 00:20
Temp : 20
Max Temp : 24 - YES
Accel : MAX
Decel : 5 (mins)
Prog Recall:
Setup: 1
Switch steps, save, or exit program mode
ENTER
when done.
STEP1
x G
STEP2
SAVE
YES
EXIT
Step #1
" appears.
SAVE
soft key, then press the
STEP2
soft key.
STEP1
STEP2
x G
SAVE
EXIT
Step #1
STEP2
soft key.
EXIT
soft key to exit
PN J3HC-IM-5AD

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents