Block Diagram; Operational Check - Agilent Technologies M9302A Service Manual

Pxi local oscillator 3 ghz to 10 ghz
Hide thumbs Also See for M9302A:
Table of Contents

Advertisement

High-Level Diagnostic Tools, Processes and References

Block Diagram

Operational Check

This operational check procedure is used for high-level troubleshooting – a rough diagnostic to ascer-
tain if the module is functioning properly. You do not have to conduct the operational check if you
conduct the
Requirements for Operational Check
The Agilent M9302A PXI Local Oscillator operational check measures the LO OUT signal and uses that as a ref-
erence to compare the signal paths to the 10 MHz REF OUT and the two 100 MHz REF OUT connectors. The fol-
lowing process demonstrates that all associated connectors, cables and circuitry are operational.
Required Hardware
To demonstrate that the module works properly requires external equipment. You will not be using the module
interconnect cables, so some high-quality flexible 3.5 mm cables will be required, plus adaptors to connect the
PSG (precision signal generator) and power sensors to 3.5 mm connectors. Please refer to the following table for
recommended hardware.
Hardware
Agilent N1913A
Agilent N8485A, Option 100
Agilent 11730A
Agilent 11667B
Agilent N9020A-526
18
Performance Verification Tests
Description
Single Channel Power Meter
10 MHz to 26.5 GHz Power Sensor
Power Sensor Cable
Power Splitter
Signal Analyzer (optional)
(see page 21).
Agilent M9302A PXI Local Oscillator Service Guide