b. Lower the side shield so the engagement tabs drop into the slots on the module's printed circuit
board.
c. Back the side shield away from the front panel to align the screw holes.
d. Install the two screws to secure the side shield to the module.
8. Install the replacement module into the chassis.
9. Power up the chassis.
a. If you are using a remote controller, power up the computer. (If you are using an embedded con-
troller, skip to Step 10.)
10. Program the replacement module with the serial number from the defective module.
a. If you don't already have the Agilent M9392A Serial Number Update Utility, download it from
www.agilent.com/find/M9392A
Examples), and install it on your computer or embedded controller.
b. Launch the Agilent M9392A Serial Update Utility (launch from the Start menu program group "Agi-
lent Utilities") and follow the embedded instructions for programming the serial number.