Agilent Technologies M9302A Service Manual page 29

Pxi local oscillator 3 ghz to 10 ghz
Hide thumbs Also See for M9302A:
Table of Contents

Advertisement

7. Configure the signal analyzer as shown in Table 3.
Table 3
Parameter
Center Frequency
Span
RBW
Averages
Reference Level
8. Configure the reference power meter as shown in Table 4.
Table 4
Parameter
Filter/Averaging
Filter Mode/Meas Avg
Resolution (display resolution)
Relative (power measurement)
Measurement (record your findings on the Test Record Card – see page 36)
1. Determine the signal analyzer flatness correction. For each LO Frequency in Table 5:
Table 5 – Test Points
LO Frequency
3.0 GHz
3.5 GHz
4.0 GHz
4.5 GHz
5.0 GHz
5.5 GHz
6.0 GHz
6.5 GHz
7.0 GHz
7.5 GHz
8.0 GHz
8.5 GHz
9.0 GHz
9.5 GHz
10.0 GHz
a. Set the signal source frequency.
b. Set the signal source power to 0 dBm.
Agilent M9302A PXI Local Oscillator Service Guide
Value
3.0 GHz
1 kHz
100 kHz
5
20 dBm
Value
On
Auto
4
Checked
Performance Verification Tests
29