Agilent Technologies M9302A Service Manual page 23

Pxi local oscillator 3 ghz to 10 ghz
Hide thumbs Also See for M9302A:
Table of Contents

Advertisement

Specification
Table 1
Power
+16 dBm ± 2 dBm
0 dBm ± 3 dBm
0 dBm ± 3 dBm
0 dBm ± 3 dBm
Equipment
Table 2
Equipment
Power Meter
Power Sensor
Adaptor, 3.5 mm (m) to 3.5 mm (f)
(for LO OUT port)
Adaptor, SMA (f) to SMB (m)
(for REF 1 OUT and REF 2 OUT ports)
Adaptor, SMB (f) to SMB (f)
(for REF 1 OUT and REF 2 OUT ports)
Test Configuration
Figure 1
Test Procedure
Test Setup
1. Turn on the power meter (with power sensor connected) and allow for warm up of 90 minutes.
2. Install the M9302A into the PXI chassis.
3. Turn on the PXI chassis and allow it to warm up for 15 minutes.
4. Perform a power meter zero and calibration on the power sensor.
5. Open the M9302A soft front panel (SFP) and set the LO Frequency to 3.0 GHz (default value).
6. Connect the power sensor to the LO OUT connector on the M9302A (see Figure 1).
7. Configure the reference power meter as shown in Table 3.
Agilent M9302A PXI Local Oscillator Service Guide
Conditions
LO Port, 3 to 10 GHz
REF OUT Port (10 MHz)
REF1 OUT Port (100 MHz)
REF2 OUT Port (100 MHz)
Supported Models
Agilent N1914A; N1912A; E4419B
Agilent N8485A
83059C
1250-0674
1250-0672
Source
M9302A Data Sheet
Figure 2
Performance Verification Tests
23