Performance Verification Tests
Table 3
Parameter
Filter/Averaging
Filter Mode/Meas Avg
Display Resolution
Measurement (record your findings on the Test Record Card – see page 36)
1. Measure the LO OUT port. For each LO Frequency in Table 4:
Table 4 – Test Points
LO Frequency
3.0 GHz
3.5 GHz
4.0 GHz
4.5 GHz
5.0 GHz
5.5 GHz
6.0 GHz
6.5 GHz
7.0 GHz
7.5 GHz
8.0 GHz
8.5 GHz
9.0 GHz
9.5 GHz
10.0 GHz
a. Set the M9302A frequency.
b. Set the power meter measurement frequency.
c. Record the power reading from the power meter.
2. Disconnect the power sensor from the LO OUT port.
3. Connect the SMA to SMB adaptor to the power sensor and connect the power sensor to the 10 MHz REF
OUT port using the SMB (m) to SMB (m) adaptor. (see Figure 2)
4. Set the power meter measurement frequency to 10 MHz.
5. Record the power reading from the power meter for the 10 MHz REF OUT port.
6. Disconnect the power sensor from the 10 MHz REF OUT port and connect it to the 100 MHz REF 1 OUT
port. (see Figure 2)
7. Set the power meter measurement frequency to 100 MHz.
8. Record the power reading from the power meter for the 100 MHz REF 1 OUT port.
24
Value
On
Auto
4
Agilent M9302A PXI Local Oscillator Service Guide