Getting Started
a. Install the module into the slot
of the chassis by placing the
module card edges into the
front module guides (top and
bottom).
b. Slide the module to the rear of
the chassis and ensure that the
injector/ejector handle is
pushed down in the unlatched
(downward) position.
c. Slide the module completely
into the chassis. When you
begin to feel resistance, pull up
on the injector/ejector handle to
fully inject the module into the
chassis.
7. Secure the front panel to the chassis using the two module front-panel mounting screws. Performance may
suffer if the screws are not tightened properly.
8. Verify that the PXI chassis fans are operable and free of dust and other contaminants that may restrict airflow.
9. Install filler panels and slot blockers after installing the module. Missing filler panels or slot blockers may dis-
rupt air circulation in the chassis.
10. Use the Cabling Diagram plus the Cable and Module Table on the next page to attach the cables to the instru-
ment. The torque specification for SMA connectors is 8 Lb-In (0.904 Nm).
11. If you are using a PCIe Cable Interface, such as the Agilent M9021, connect the Cable Interface in the chassis
to the PC host per the instructions that came with the Cable Interface.
12. Power up the PXI chassis.
13. Reboot the PC host.