RM640
Menu No. Function
C7.1.6
Software
Revision
C7.1.7
Electronic
Serial No.
C7.1.8
Production Date This the date that the manufacturer finished
C7.1.9
Calibration Date This the date that the manufacturer
C7.2 Security
C7.2 Security
C7.2 Security
C7.2 Security
C7.2.1
Login
C7.2.2
Change
Password
C7.2.3
Reset
Passwords
C7.2.4
Unlock
Extended Range
C7.2.5
Unlock SIL
C7.3 Errors
C7.3 Errors
C7.3 Errors
C7.3 Errors
C7.3.1
Message View
C7.3.2 Error-Mapping
C7.3.2.1
Sensor:
Information
C7.5 Units
C7.5.1
Length
34-VF-25-131 iss.1 GLO Feb 2019 US
Function description
This is the software revision number. This
number agrees with NAMUR NE 53
Guidelines.
This is the hardware fabrication number.
This number agrees with NAMUR NE 53
Guidelines.
the device. The date format is: Year-Month-
Day.
calibrated the device. The date format is:
Year-Month-Day.
Enter the appropriate password here to
change settings. If you do not enter the
password, you can only change settings for
the "user" access level. For more data and
the default passwords for the "operator"
and "expert" access levels, refer to
Protection of the device settings (access
levels)
on page 68.
This changes the password for the
"operator" and "expert" access levels. For
the procedure to change the password,
Protection of the device settings
refer to
(access levels)
on page 68 (How to change
the password).
This is a special password that makes your
passwords for the "operator" and "expert"
access levels go back to the default
passwords. This password is supplied on
request by the after-sales service
department.
This menu item can only be unlocked at the
factory.
This menu item can only be unlocked at the
factory.
A log of device errors. Scroll down the list
and push [> > > > ] to show the error details. The
error will have a letter code ("F", "S", "M",
"C" and "I") that agrees with NAMUR NE 107
Guidelines.
This permits you to change the error code
that is given to an incident.
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
setting: User
setting:
setting:
setting:
The length unit shown in Normal mode. If
you set this menu item to "Cst." (custom
length unit), enter values in menu items
C7.5.2.1 thru C7.5.2.3.
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
setting: User
setting:
setting:
setting:
www.honeywellprocess.com
OPERATION
Selection list
Read only
Read only
Read only
Read only
4-digit hexadecimal
password
4-digit hexadecimal
password
Read only
None, Information (I),
Maintenance Request
(M), Out of Specification
(S), Function Check (C),
Failure (F)
m, cm, mm, ft, in, Cst.
6
Default
Refer to
"Function
description"
Refer to
"Function
description"
Information
m
97