Replacement Of The Smartline® Rm60 Signal Converter With The Rm640 Signal Convert- Er - Honeywell RM640 Handbook

Non-contact radar level meter - 24 ghz radar (fmcw) level transmitter for solids from granulates to rocks
Table of Contents

Advertisement

SERVICE
7
7.3 Replacement of the SmartLine® RM60 signal converter with the RM640
signal converter
INFORMATION!
Complete the 4 procedures that follow in numerical sequence.
Equipment needed
Figure 7-2: Equipment needed
1 5 mm Allen wrench (not supplied)
2 SmartLine® RM60 radar level transmitter
3 RM640 signal converter (without process connection and antenna). It is necessary to send an order for the
OPTIWAVE 6400 signal converter with an adaptor for the OPTIWAVE 6300 process connection attached. For the spare
parts order code, refer to
of the SmartLine® RM60).
4 Handbooks for all devices. You can download this document free of charge from the website.
CAUTION!
SmartLine
SmartLine
SmartLine
SmartLine
Record the SmartLine
with your order for the RM640 signal converter.
Make sure that you also record the device settings for the SmartLine
basic configuration (tank height, blocking distance etc.), output, application, display and
strapping table data. You can find this data in Supervisor mode.
Procedure 1: Record the device serial number (SmartLine® RM60 radar level
transmitter) and send the order for the signal converter
• Find the serial number on the SmartLine® RM60 nameplate.
• Record this number.
• Create an order code for the RM640 signal converter. Make sure that the RM640 signal
converter has the adaptor option for the SmartLine® RM60 flange system.
124
Spare parts
on page 170 (Table: Replacement signal converter for mechanical sub-assembly
RM60:
RM60
®
RM60
RM60
RM60 serial number on the device nameplate. Send the serial number
®
www.honeywellprocess.com
RM640
®
RM60. This data includes
34-VF-25-131 iss.1 GLO Feb 2019 US