Calibration; Cal:pres:poin - GE PACE Series Technical Manual

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

PACE Series SCPI Manual
The CALibration subsystem enables the calibration of the transducers and the rate control
system, refer to the user manual for further details.
Applicability:
Command Syntax
n/a
Parameter:
Short form:
Function:
Defaults:
Query Syntax
:CALibration[x]:[PRESsure]:POINts?
where: x (module) = 1 or 2 (default - 1).
Short form:
Function:
Response:
Description
Valid only when calibration is enabled, this queries the number of calibration points.
Example

CALibration2:

Tx> :
Rx> :CAL2:
PRES:
K0472 Revision A
CALibration

:CAL:PRES:POIN

PACE1000, PACE5000, PACE6000
CAL:POIN?
Gets the number of calibration points
3
POIN
PRESsure:
ts?
POIN
3
4-3