3.1

Output queue

The output queue is a text readable data queue that is read through the IEEE 488 talk
command. The queue is cleared by reading all elements in it or by the *CLS command.
Every time a query has been successfully completed, the response, in a text readable format
is placed at the end of the output queue. If the MAV bit in the "Status Byte" was previously
cleared it will be set. The output queue can contain up to 256 characters. If there is not
enough space in the output queue for a new message, the error -350, "Queue overflow" will
be placed into the error queue and the most recent output message will be lost.
K0472 Revision A
3 - 3