GE PACE Series Technical Manual page 117

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

PACE Series SCPI Manual
:SOUR:PRES:LEV:IMM:AMPL:VENT
Applicability:
Command Syntax
:SOURce[x][:PRESsure][:LEVel][:IMMediate][:AMPLitude]:VENT
<number>
where x = 1 or 2 is the module number (default - 1)
Parameter:
Short form:
Function:
Default:
Query Syntax
:SOURce[x][:PRESsure][:LEVel][:IMMediate][:AMPLitude]:VENT?
where x = 1 or 2 is the module number (default - 1)
Short form:
Function:
Response:
Description
This command vents the user system; the command should be queried to get the status of
the vent.
Examples
For an instrument switched on without any vent, a query:
TX> :SOUR:PRES:LEV:IMM:AMPL:VENT?
RX> :SOUR:PRES:LEV:IMM:AMPL:VENT 0 (a return of '0' indicates vent not in progress).
A vent can be started by sending:
TX> :SOUR:PRES:LEV:IMM:AMPL:VENT 1
K0472 Revision A
PACE5000, PACE6000
<integer>
0 - abort vent
1 - start vent
SOUR:VENT <integer>
Vents the user system.
0
:SOUR:VENT?
Query status of vent.
0 - vent OK
1 - vent in progress
2 - vent completed
4-89