Sens:pres:corr:head:stat - GE PACE Series Technical Manual

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

4 Command and Query Summary

:SENS:PRES:CORR:HEAD:STAT

Applicability:
Command Syntax
:SENSe[x][:PRESsure]:CORRection:HEAD:STATe <Boolean>
where x = 1 or 2 is the module number (default - 1)
Parameter
Short form::
Function:
Default:
Query Syntax
:SENSe[x][:PRESsure]:CORRection:HEAD:STATe?
where x = 1 or 2 is the module number (default - 1)
Short form:
Function:
Response:
Description
This command enables or disables the head correction compensation.
Example
TX> :SENS:PRES:CORR:HEAD:STATe?
RX> :SENS:PRES:CORR:HEAD:STAT 0
Head correction off
It can be turned on and off by:
TX> :SENS:PRES:CORR:HEAD:STATe on
TX> :SENS:PRES:CORR:HEAD:STATe off
Or by:
TX> :SENS:PRES:CORR:HEAD:STATe 1
TX> :SENS:PRES:CORR:HEAD:STATe 0
K0472 Revision A
PACE1000, PACE5000, PACE6000
<boolean>
0
-
1
-
SENS:CORR:HEAD:STAT <Boolean>
Enables/disables head correction.
0
:SENS:CORR:HEAD:STAT?
Query head correction state
0 - head correction off
1 - head correction on
Disables head correction
Enables head correction