Sens:alt:inl - GE PACE Series Technical Manual

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

PACE Series SCPI Manual
Applicability:
Command Syntax
:SENSe[x][:ALTitude]:INLimits <number>
where x = 1 or 2 is the module number (default - 1)
Parameter:
Short form:
Function:
Default:
Query Syntax
:SENSe[x][:ALTitude]:INLimits?
where x = 1 or 2 is the module number (default - 1)
Short form:
Function:
Response:
Description
Sets the in-limits value.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.
K0472 Revision A

:SENS:ALT:INL

PACE6000
<decimal> in limits value as % full-scale
:SENS:INL <number>
The instrument has an in-limits indicator. This can generate a
service request when the pressure is within limits for a set time period.
0.01 % full-scale
minimum
0 % full-scale
maximum 100 % full-scale
:SENS:INL?
Query in-limits value
Decimal representing in-limits value as % full-scale.