Calibration Instructions - GE Druck UPS-III User Manual

Loop calibrator
Hide thumbs Also See for Druck UPS-III:
Table of Contents

Advertisement

0.561.8187

Calibration Instructions

General
The instrument is supplied by the manufacturer, complete
with calibration certificate(s). A calibration period of 12
months is recommended. The actual calibration interval
depends on instrument usage and the total measurement
uncertainty acceptable for the specified application.
The UPS-III is a very precise measuring instrument and the test
equipment and conditions of test must be suitable for the type
of work. The calibration check and calibration adjustment
should be carried out in a controlled environment by a
calibration technician*.
The manufacturer offers a comprehensive and, if required,
UKAS accredited calibration service.
*
A calibration technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the calibration
work on this equipment.
Calibration Equipment
The following tables give the accuracy requirements for the
calibration equipment and the UPS-III.
Calibration requires a stable temperature of 21° ±1°C (70° ±2°F).
UPS-III measure mode
Applied
Permitted UPS-III error
mA
(mA)
0
0.002
4
0.002
12
0.002
20
0.002
www.
Table 1
mA measure
Calibrator error
(mA)
0
0.00014
0.00030
0.00046
13
K0317 Issue No. 3
.com
information@itm.

Advertisement

Table of Contents
loading

Table of Contents