Sens:pres:res - GE PACE Series Technical Manual

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

4 Command and Query Summary
Applicability:
Command Syntax
:SENSe[x][:PRESsure]:RESolution <string>
where x = 1 or 2 is the module number (default - 1)
Parameter:
Short form:
Function:
Query Syntax
:SENSe[x][:PRESsure]:RESolution?
where x = 1 or 2 is the module number (default - 1)
Short form:
Function:
Response:
Description
This command selects the resolution to be used for showing the pressure reading in the front
panel display.
TX> :SENS:PRES:RES?
RX> :SENS:PRES:RES 6
Example 1
The resolution can be changed by the command, such as:
TX> :SENS:PRES:RES 4
Query again confirms that the resolution has been changed to '4' :
Example 2
TX> :SENS:PRES:RES?
RX> :SENS:PRES:RES 4
Any resolution number outside the 4 to 7 boundary generates error messages,
such as:
TX> :SENS2:PRES:RES 8
TX> :SYST:ERR?
Example 3
RX> :SYST:ERR -222,"Data out of range; Parameter 1"
TX> :SENS2:PRES:RES 3
TX> :SYST:ERR?
RX> :SYST:ERR -222,"Data out of range; Parameter 1"
K0472 Revision A

:SENS:PRES:RES

PACE1000, PACE5000, PACE6000
<integer>
resolution information
:SENS:RES
Used to select a resolution to be used.
:SENS:RES?
Asks for current resolution.
An integer from 4 to 7 representing selected resolution.