GE PACE Series Technical Manual page 46

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

4 Command and Query Summary
Description
This query returns a string representing the range, a number representing the upper full-
scale and lower full-scale.
Example for a dual module instrument, both modules with the barometric option fitted.
TX> :INST:LIM?
RX> :INST:LIM "7.00barg", 7350.0000000, -1100.0000000
TX> :INST:LIM1?
RX> :INST:LIM "7.00barg", 7350.0000000, -1100.0000000
TX> :INST:LIM2?
RX> :INST:LIM2 "20.00barg", 21000.0000000, -1100.0000000
TX> :INST:LIM3?
RX> :INST:LIM3 "2.00barg", 2100.0000000, -1100.0000000
TX> :INST:LIM4?
RX> :INST:LIM4 "BAROMETER", 1207.5000000, 825.0000000
TX> :INST:LIM5?
RX> :INST:LIM5 "3.50barg", 3675.0000000, -1100.0000000
TX> :INST:LIM6?
RX> :INST:LIM6 "20.00barg", 21000.0000000, -1100.0000000
TX> :INST:LIM7?
RX> :INST:LIM7 "10.00barg", 10500.0000000, -1100.0000000
TX> :INST:LIM8?
RX> :INST:LIM8 "BAROMETER", 1365.0000000, 38.5000000
K0472 Revision A
4-18

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pace1000Pace5000Pace6000

Table of Contents