Sour:pres:inl - GE PACE Series Technical Manual

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

4 Command and Query Summary
Applicability:
Command Syntax
:SOURce[x][:PRESsure]:INLimits <number>
where x = 1 or 2 is the module number (default - 1)
Parameter:
Short form:
Function:
Default:
Query Syntax
:SOURce[x][:PRESsure]:INLimits?
where x = 1 or 2 is the module number (default - 1)
Short form:
Function:
Response:
Description
Sets the in-limits value.
Example
TX> :SOUR:PRES:INL?
RX> :SOUR:PRES:INL 0.0200000
The current in-limits set-point is 0.02% full-scale
Can be set to:
TX> :SOUR:PRES:INL 0.01
TX> :SOUR:PRES:INL?
RX> :SOUR:PRES:INL 0.0100000
K0472 Revision A

:SOUR:PRES:INL

PACE5000, PACE6000
<decimal> in limits value as % full-scale
:SOUR:INL <number>
The controller has an in-limits set-point indicator. This can generate a
service request when the pressure is within limits for a set time period.
0.01 % full-scale
minimum
0% full-scale
maximum 100% full-scale
:SOUR:INL?
Query in-limits value
Decimal representing in-limits value as % full-scale.
4-86