PACE Series SCPI Manual
Command Syntax

*OPC

Parameter:
Short form:
Function:
Default:
Note:
Query Syntax
*OPC?
Short form:
Function:
Response:
Note:
Description
This command sets the OPC bit in the standard event register. The query returns '1' when all
operations have been completed.
K0472 Revision A
*OPC
None
*OPC
When all operations have been completed, event 800 is written to the
error/event queue (if enabled by STAT:QUE:ENAB). At the same time, this
command allows the OPC bit in the standard event register to be set. This
can be used to trigger a serial poll, see *ESR for details. If there are no
pending operations these actions are performed immediately. This
command should be sent each time the OPC bit is allowed to be set.
-
It is intended that the *OPC command should be the last command in a
program message.
*OPC?
When all operations have been completed a '1' is written to the output
queue this can be used to trigger a serial poll, see *SRE for details.
If there are no pending operations a '1' is written to the queue
immediately.
It is intended that the *OPC command should be the last command in a
program message.
4-135