Input; Inp:log - GE PACE Series Technical Manual

Pressure automated calibration equipment, scpi remote communications manual - k0472 revision a
Table of Contents

Advertisement

4 Command and Query Summary
The INPut subsystem shows the state of the logical inputs.
Applicability:
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:INPut[x]:LOGic?
where: x (module) = 1 or 2 (default - 1),
Short form:
Function:
Response:
Description
This query returns the state of the switch input within the module and the pressure at time of
switching operations.
Example

:INP:LOG?

:INP:LOG 0, 0.8321209
Current logic OFF, pressure was 0.8321209 in current pressure units when the module was
switched to OFF condition.
K0472 Revision A

INPut

:INP:LOG
PACE5000, PACE6000
:INP:LOG?
Asks for state of switch input within the controller module.
first parameter
second parameter
-
0 = OFF, 1 = ON
-
measured pressure at the time of
switching (snapshot in current pressure
units).
4-12

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pace1000Pace5000Pace6000

Table of Contents