Modbus communication
A.9 Modbus holding registers tables
Modbus
Data type /
register
Size in bytes
6198
Unsigned / 2
6190
String / 16
10864
String / 32
A.9.4.6

Local display

General settings
Table A-72
General settings
Modbus
Data type /
register
Size in bytes
9002
Unsigned / 2
9098
Unsigned / 2
9094
Unsigned / 2
9095
Unsigned / 2
9096
Unsigned / 2
9097
Unsigned / 2
9040
Unsigned / 2
268
Parameter
Description
Set time
Command to set the de‐
vice's current time by
using the separate pa‐
rameters above.
Current time
Current time in ISO
8610 format
(YYYY-MM-DD
hh:mm).
Current time
Current time in ISO
high resolution
8610 format
(YYYY-MM-DD
hh:mm).
Parameter
Description
Language
Language.
Backlight
Duration of backlight
activity after last key
press.
Inactivity time‐
Duration of display ac‐
out
tivity after last key
press.
Auto logout
Defines the need to en‐
ter PIN after leaving the
menu and returning to
any top level view.
Help appear‐
Duration until help text
ance
appears.
Enable local ac‐
Controls local access.
cess control
Status icons
Type of alarm icons
shown in local display.
Default value
Value range /
[unit]
Setting options
(unit register)
-
● 0: Cancel
● 1: Set time
-
-
-
-
Default value
Value range /
[unit]
Setting options
(unit register)
0
● 0: English
● 1: German
30
0 - 240
[s]
10
0 (always on) - 60
[min]
1
● 0: Enter menu without
PIN for 10 min.
● 1: Automatic logout.
3
0 - 20
[s]
1
● 0: Disabled. Allows
users to modify
settings without PIN.
Expert access is
excluded.
● 1: Enabled.
Order-specif‐
● 0: Standard
ic.
(Siemens)
● 1: NAMUR
Operating Instructions, 08/2018, A5E37367739-AD
Access
level
Write on‐
ly
Read
only
Read
only
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
SITRANS FST030 (Modbus)

Hide quick links: