Data Logging - Siemens SITRANS F Series Operating Instructions Manual

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Modbus communication
A.9 Modbus holding registers tables
Modbus
Data type /
register
Size in bytes
6182
Unsigned / 2
6099
Unsigned / 2
A.9.9.2

Data logging

Table A-117 Data logging
Modbus
Data type /
register
Size in bytes
11400
Unsigned / 2
11401
Unsigned / 2
11402
Unsigned / 4
318
Parameter
Description
Backup
Command to overwrite
the existing backup da‐
ta sets.
This ensures that the
current sensor serial
number and transmitter
serial numbers are up‐
dated in the backup da‐
taset. This enables the
automatic backup.
Restore
Command to restore
backup data. Coding is
bitwise.
Parameter
Description
Enable / disable Controls data logging.
Data logging
Controls data logging
mode
mode.
Logging interval Defines interval for log‐
ging. Granularity is 10
ms.
Default value
Value range /
[unit]
Setting options
(unit register)
-
● 1: Overwrite existing
backup files
-
● Bit 0: Restore sensor
setup parameters
● Bit 1: Restore sensor
application setup
parameters
● Bit 2: Restore
transmitter
application setup
parameters
● Bit 3: Restore
customer transmitter
configuration setup
parameters
● Bit 4: Restore
totalizers
● Bit 5: Restore system
setup
● Bit 6: Restore
customer sensor
configuration setup
parameters
Default value
Value range /
[unit]
Setting options
(unit register)
0
● 0: Disabled
● 1: Enabled
0
● 0: Log instantaneous
● 1: Log average value
500
1 - 8640000
[x 10 ms]
Operating Instructions, 08/2018, A5E37367739-AD
Access
level
Write on‐
ly
Write on‐
ly
Access
level
Read /
write
Read /
write
Read /
write
SITRANS FST030 (Modbus)