Siemens SITRANS F Series Operating Instructions Manual page 212

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Modbus communication
A.9 Modbus holding registers tables
Modbus
Data type /
register
Size in bytes
10040
Float / 4
10042
Float / 4
10068
Float / 4
10070
Float / 4
8908
Float / 4
8939
Unsigned / 2
8940
Float / 4
7149
Float / 4
212
Parameter
Description
Upper range
Upper process viscosi‐
value viscosity
ty value that is mapped
to the upper nominal
range boundary selec‐
ted by Frequency value
high (8935). Applicable
if Process value (8904)
= 21 or 29.
Lower range
Lower process viscosi‐
value viscosity
ty value that is mapped
to the lower nominal
range boundary selec‐
ted by Frequency value
low (8937). Applicable
if Process value (8904)
= 21 or 29.
Upper range
Upper rate of change
value ROC
value that is mapped to
the upper nominal
range boundary selec‐
ted by Frequency value
high (8935). Applicable
if Process value (8904)
= 36.
Lower range
Lower rate of change
value ROC
value that is mapped to
the lower nominal
range boundary selec‐
ted by Frequency value
low (8937). Applicable
if Process value (8904)
= 36.
Damping value
Time constant of damp‐
ing filter for frequency
output signal.
Fail-safe behav‐
Frequency output reac‐
ior
tion in case of a fault.
Fail-safe value
Output value in case of
a fault and when Fail-
safe behavior (8939) is
configured to 4.
Forced value
Forced value when
forcing is active.
Default value
Value range /
[units]
Setting options
(units register)
1.0
-
[m²/s]
(7524)
0.0
-
[m²/s]
(7524)
1.0
-
[m/s²]
(8564)
0.0
-
[m/s²]
(8564)
0.0
0.0 - 100.0
[s]
2
● 0: Frequency value
low
● 1: Frequency value
high
● 2: Last valid value
● 3: Disabled
● 4: Custom value
0.0
0.0 - 12500.0
[Hz]
1.0
0.0 - 12500.0
[Hz]
Operating Instructions, 08/2018, A5E37367739-AD
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
SITRANS FST030 (Modbus)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fst030

Table of Contents