Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
ID/
Diagnostic
Icons
162
Density
163
Medium temperature
166
Standard volume flow
167
Totalizer 1
168
Totalizer 2
169
Totalizer 3
170
Loop current
172
Transmitter
173
Sensor
110
Action
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Loop current simulated. Disable 'Simulation' be‐
fore returning to normal operation
FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
FW invalid. A component does not have the ex‐
pected FW version. Start a product firmware
update to update the component version or re‐
place the component.
Comment
Only hydrocarbon
Only HART devices
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD