Siemens SITRANS F Series Operating Instructions Manual page 109

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

ID/
Diagnostic
Icons
147
Totalizer 3
148
Transmitter temperature
149
Transmitter temperature
150
Sensor signal disrupted
151
SensorFlash
152
SensorFlash
159
Internal error
160
Mass flow
161
Volume flow
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD
Action
Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Lower alarm limit'
Transmitter temperature above alarm limit. Re‐
locate the device if possible or lower the ambi‐
ent temperature enough to cool the device. In‐
spect the device for heat-related damage.
Transmitter temperature below alarm limit. Re‐
locate the device if possible or raise the ambient
temperature enough to heat the device. Inspect
the device for cold-related damage.
Turn off the power. Unplug and reconnect the
sensor cable. Restore power. If the error still
exists, contact your local Siemens representa‐
tive.
Backup disabled. Another SensorFlash was in‐
serted. To acknowledge please copy the con‐
figuration from the SensorFlash to the device,
remove this Sensorflash and insert the original
SensorFlash.
Backup disabled. Another SensorFlash was in‐
serted. To acknowledge please copy the con‐
figuration from the SensorFlash to the device,
remove this Sensorflash and insert the original
SensorFlash.
Internal error in transmitter. Turn off the power,
wait 5 seconds and turn on the power again. If
the error still exists, contact your local Siemens
representative.
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Value simulated. Disable 'Simulation' before re‐
turning to normal operation
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
Comment
109