Modbus communication
A.9 Modbus holding registers tables
Path 1 settings
Table A-21
Path 1 settings
Modbus
Data type /
register
Size in bytes
1450
Float / 4
1458
Unsigned / 2
1465
Float / 4
1459
Float / 4
Path 2 settings
Table A-22
Path 2 settings
Modbus
Data type /
register
Size in bytes
1500
Float / 4
1508
Unsigned / 2
170
Parameter
Description
Sensor frequen‐
Desired sensor frequency for
cy
path 1.
Path geometry
The number of crossings the sig‐
nal makes through the pipe (Re‐
flect or direct geometry for path
1).
E.g.:
Z path = 1
V path = 2
W path = 4
Clamp-on spac‐
Spacing (LTN) between clamp-
ing
on sensors for path 1.
Path weight
Fixed weighting factor associ‐
ated with path 1.
For clamp-on systems the instal‐
led paths weightings should be
identical but not zero.
For inline systems the sum of
the installed paths weightings
should be 1.0.
Parameter
Description
Sensor frequen‐
Desired sensor frequency for
cy
path 2.
Path geometry
The number of crossings the sig‐
nal makes through the pipe (Re‐
flect or direct geometry for path
2).
E.g.:
Z path = 1
V path = 2
W path = 4
Default value
Value range /
[units]
Setting options
(units register)
1100000
80000.0 -
[Hz]
4000000.0
1
1 - 10
0.0
-100.0 - 100.0
[m]
(8556)
1.0
0.01 - 2.0
Default value
Value range /
[units]
Setting options
(units register)
1100000
80000.0 -
[Hz]
4000000.0
1
1 - 10
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
(Expert)
Access
level
Read /
write
Read /
write