ILLUSTRATION 5-18
IMAGE MONITOR SCREEN

2-4Video Presentation

2-5Fluoro Time

om 5184516-100 Rev. 5
The Image Monitor will display the following screen:
These selections will change image presentation head/foot and left/right.
Total fluoro time is displayed.
The 5 minute fluoro timer can be rest by pressing the button. An audible alarm will
sound at the completion of 5 minutes. You may continue to fluoro with the audible
tone for another 5 minutes. However at the end of the second 5 minute period the
time must be reset otherwise fluoro will be discontinued until this is done.
Total patient dose and Dose area product is displayed in the upper right corner of
the screen.
5-18