Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
DISPLAY TEXT
8.4
Data print out of actual program parameter:
Comment: With the PRINT key you get to the printer menü.
Equipment data
Program parameter
Current data
Configuration
Print out
every nnn minutes?
Comment: When printed out as recommended a complete test documentation is available.
8.5
Change speed of blower during operation:
Depending on lamp ageing and the kind of specimen it can be necessary to change the fan speed of
ALPHA und HIGH ENERGY during operation. After starting the unit press the DATA key until the following
menu appears:
Blower:
hh:mm
Comment: Input a number between „500" and „2500".
8.6
Stop a running program:
A running test program can be stopped by pressing the STOP key.
Program interrupted
Continue program:
---> START
Comment: To continue the program press START. The test program continues at the point of
interruption.
ALPHA aund HIGH
Safety instructions
ENERGY features
Function
Program example
Page 54
= 1
= 2
= 3
= 4
1800 rpm
dd.mm.yy
Unit set-up
Setting up tests
INPUT/KEY PAD
3
nn
ENTER
STOP
ESC
Unit description
Options und accessory
Care and cleaning
Maintenance
ATLAS
Start-up
List of options and
accessories
Need help?