Water Quality; Water Supply - Atlas XENOTEST ALPHA Operating Manual

Table of Contents

Advertisement

ATLAS
3.2.4

Water quality

The water must be of high-grade purity
to prevent deposits.
The use of ion exchange cartridges alo-
ne is not sufficient.
Regarding the water purity we refer to the mini-
mum requirements of some important standards,
such as DIN 53387 „Artifical light and weather fast-
ness of plastics":
1.
Electrical conductivy: < 5 µS/cm
2.
PH value: 6-8
3.
content of solid bodies < 1 ppm measured
as silicon dioxide.
ALPHA and HIGH
Safety instructions
ENERGY features
Function
Program example
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
3.2.5
To ensure the functioning of the water
circuit and the humidification system,
the apparatus must be aligned by
means of its height adjustable feet.
By supply line:
fill the water tank (fig.9) with a pressure-resistant
hose to the water inlet of the tank (inner diameter
13mm, 3bar pressure).
without supply line:
fill the water tank manually.
Unit set-up
Unit description
Setting up tests
Care and cleaning

Water supply

Options und accessory
List of options and
Maintenance
Start-up
accessories
Page 13