5.3.1.2 Turning the Degasser on and off in Chromeleon

If the pump is operated from Chromeleon, select one of the following alternatives to turn the
degasser on and off:
Directly from the Panel Tabset (→ page 33)
Directly from the Commands dialog box for the pump (→ page 34)
Automatically with a program file (PGM) for the pump (→ page 34)
If the Solvent Rack is connected to an SD(N), RS, or BM pump or to an NCS-3500RS or
NCP-3200RS, verify in the Properties dialog for the device that Degasser Control is set to
External (with SD(N), RS, and BM pumps on the Devices page; with an NCS/NCP on the
Pumps page).
To turn the degasser on or off from the Panel Tabset
1. In Chromeleon, click the View menu and Default Panel Tabset or click the
corresponding icon on the toolbar
Chromeleon creates centralized control panels, called panel tabsets, for all timebases
available on the Chromeleon server. A panel tabset provides control panels for the
individual instruments in a timebase and, in addition, one or more panels for performing
system-wide functions, for example, creating and running sequences. For more
information about panel tabsets, see the Chromeleon Help.
2. On the panel tabset for your timebase, click the pump page (→ Fig. 11) or NCS/NCP
page.
Operating Instructions
, and then connect to the Chromeleon server.
Fig. 11: Pump control panel on the panel tabset
UltiMate 3000 Series:
SRD-3x00 Solvent Racks
Page 33