♦ Some capillaries of the RS pumps and some Viper system capillaries are made of
®
MP35N
, a nickel-cobalt based alloy. Individuals with sensitivity to nickel/cobalt may
show an allergic reaction from skin contact.
♦ Always wear safety glasses when handling fused silica tubing, for example, during
installation or when cutting capillaries to the length.
•
For safety reasons, you shall not lift the SRD with one or more solvent reservoirs being in
the tray.
•
The enclosure must be opened only by Thermo Fisher Scientific service personnel.
Replace faulty communication cables.
•
•
Replace faulty power cords. Never use a power cord other than the power cords provided
for the device.
Use only the original spare parts and accessories authorized for the device by Thermo
•
Fisher Scientific.
•
When operating the HPLC system, always set a lower pressure limit for the pump. This
prevents damage resulting from leakage or from running the pump dry over a longer
period. Activate solvent reservoir level monitoring (→ Pump manual).
•
Use only standard HPLC grade solvents (MS grade if the SRD is operated together with
an NCS-3500RS or NCP-3200RS) and buffers that are compatible with all parts that may
be exposed to solvents (→ page 49).
The front panel tilts upward. When lifting or moving the SRD, always lift by the bottom
•
or sides of the instrument. This is to avoid damage to the instrument.
•
The open front panel door is not designed to carry weight. Therefore, you shall not place
any objects on the open front panel door.
•
If a leak occurs, turn off the SRD and remedy the situation immediately.
•
Before interrupting operation for several days or more, observe the precautions on
page 36.
•
Do not use the SRD in ways other than those described in these operating instructions.
•
Keep the operating instructions near the device to be available for quick reference.
Operating Instructions
UltiMate 3000 Series:
SRD-3x00 Solvent Racks
Page 5