Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 24

Solvent racks
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
SRD-3x00 Solvent Racks
Important: To prevent the device from falling, grasp the device by the sides,
Important: Afin d'empêcher l'instrument de tomber, saisissez-la par les côtés.
3. Remove the foam spacers, and then remove the polythene packaging.
4. Before connecting the SRD to the power source, wait approximately four hours to
allow the instrument to come to room temperature and to allow any condensation that
might have occurred during shipping to evaporate. After four hours, check the SRD; if
condensation still exists, allow the instrument to continue to warm up (without
connecting it to the power source) until the condensation is completely gone.
Page 18
and then lift the unit together with the foam spacers out of the
shipping container. Do not lift the device by the foam spacers or
the front panel.
Ne soulevez l'instrument à l'aide du matériau d'emballage ou par la
porte du panneau avant.
Operating Instructions